The reasons for the oxygen concentration of the oxygen concentrator becomes low
There are several reasons why the oxygen concentrator's concentration may become low:
Clogged air filters: the filters in the oxygen concentrator can become clogged with dust, dirt, or other particles over time, which can reduce the flow of air into the machine and decrease the concentration of oxygen produced.
Oxygen flow rate adjustment is beyond the rated flow: The oxygen generator cannot run higher than the rated flow rate. If the rated flow rate of the oxygen generator is 5L, if the user adjusts it beyond the 5L scale, the oxygen concentration will decrease.
The atomization function is turned on while inhaling oxygen: During the atomization process, the air participates in the atomization, which reduces the oxygen raw material air and reduces the oxygen concentration.
Molecular sieve failure: Molecular sieve is the core component of the oxygen generator and is easily affected by the environment. If the oxygen production capacity decreases, you need to contact the manufacturer to replace the molecular sieve.
Oxygen sensor malfunction: The oxygen concentrator relies on sensors to measure the oxygen concentration in the output. If these sensors malfunction or become inaccurate, it can result in a lower than expected oxygen concentration.
Improper maintenance: Regular maintenance is essential to ensure the proper functioning of an oxygen concentrator. If it is not cleaned or serviced regularly, it can lead to a decrease in oxygen concentration.
Aging components: Over time, the components of an oxygen concentrator may wear out or deteriorate, affecting its performance. This can result in a lower oxygen concentration.
Note: If the oxygen concentrator is not used for a long time, and the oxygen machine is exposed to the air, without any measures, the oxygen concentration will also decrease.
The treatment plan varies from person to person, please purchase according to the doctor's advice, and please refer to the product manual or video tutorial for the product usage steps.
It is important to note that if you suspect a decrease in oxygen concentration from your oxygen concentrator, it is recommended to contact the manufacturer or a qualified technician for assistance and proper diagnosis.